KKC3.Ex/8-K33 Galltec Mela United Kingdom Sales
If you are looking for a reliable supplier where you can purchase the KKC3.Ex/8-K33 humidity/temperature sensor, you are in the right place. RADWARD sells all product groups belonging to the Galltec Mela brand in the United Kingdom market, with fast delivery times and competitive prices.
Technical Specification
Humidity/temperature sensor Ex version
suitable for use in EO atmosphere;
pressure-resistant 50mbar (non-atmospheric.
conditions)
II 1/2 G Ex ia IIC T4 Ga/Gb
II 2 D Ex ia IIIC T95°C Db
Ta: -40... +80°C
Approved for use in
potentially explosive areas
according to EC type examination certificate
iBEXu 07 ATEX 1114
Use in zone 0 only together with
ZA40 mounting kit
(Please order ZA40 mounting kit separately,
further installation instructions and dimensions
according to data sheet)
Die-cast aluminum housing
Stainless steel sensor tube incl. sinter filter
MB Humidity: 0... 100% RH
MB temperature: -20... 80 °C
output: 2 x 4... 20 mA
Operating point: 13... 24 V DC intrinsically safe
sewer installation,
Dip tube length: 232 mm
-------------------------------------------------------------
----------------------------------
This device is powered by
The ones in datasheet C4.8
listed transmitter power supplies. plea
order separately.
Using another
food appliances
is done at the user's risk.
-------------------------------------------------------------
------------------------------------
Use in accordance with the data sheet
Subject to change
_____________________________________
_______________
Approval for use with
gas sterilization with
Ethylene oxide at reduced pressure up to p (abs) = 50
mbar according to
Safety declaration dated 27.09.2007
to
EC type examination certificate
iBEXu 07 ATEX 1114
This safety declaration applies to
The following
submitted with the application
operating conditions:
- Ambient pressure, 45-55°C air
- 50mbar (abs), 45-55°C air, 20% O2
- 50mbar (abs), 45-55°C air, 100% RH
120-850 mbar (abs),
45-55°C, 0-8% O2, 0-65% EO, 20-100% RH
The sensor is on the partition to
attach and
With the measuring head partially in the
Sterilization area
(zone 0). Zur
Partition wall mounting is
to use the ZA40 mounting kit.
Assembly kit please
Order separately!
EC COMPLIANCE: CE95